◆7” High Resolution Color TFT LCD
◆1-8 Channel Waveform Display
◆7 Leads ECG Waveform Display; ST Segment Analysis
◆Defibrillation and ESU Protection
◆OsyCRG diagram, Drug Dose Calculation
◆Venous Punch
◆Flexible Parameter Configuration
◆Applicable for adult, pediatric, and neonatal patients
◆Options: Thermal recorder, IBP
Product Features :
◆ECGECG Cable: 5 Leadwires ECG cable Lead section: lead I, II, III/aVR, aVL, aVF/ V
Gain(mm/mV): 1/4,1/2,1,2,4
Frequency response: Diagnosis: 0.05 – 100Hz Monitor: 0.5~40 Hz Surgery:1.0~25 Hz Sweep speed(mm/sec):6.25,12.5, 25, 50 Heart rate range: 0-350 BPM Heart rate accuracy : ± 1 % ST segment deviation analysis: -0.8~+0.8mV
ST accuracy : ±0.02mv
◆Non-Invasive Blood Pressure Measurement type : adult, pediatric, neonatal Measurement range: Systolic 4.0 – 270mmHg Diastolic 20 – 200mmHg Mean 30 – 220mmHg Accuracy : ±3mmHg or 2 %(choice max) Resolution:1mmHg
Protection: over pressure
◆Pules Oximetry SPO2 measurement range: 0 – 100 %Accuracy: ±2% at 70~100%(no specification under 70%) Resolution: 1 % pulse measurement range:30- 250BPM Accuracy: ±2 % or 1BPM (choice max)
Resolution: 1 BPM
◆Respiration Rate Measurement range: 0 – 120 BPMAccuracy: ±1 BPM or 5 %(choice max) Resolution: 1 BPM ◆CO2Measurement range: 0 % – 10.0 % (in CO2 and et CO2) Accuracy: ± 2 mm Hg(5 % Measurement value) Response time: 180 ms ◆IBPChannel number: 2chrange: -40 – +330mmHgAccuracy: ±4mmHg or ±4%(choice max)Resolution: 1mmHg◆Environment Opreation Storage Temperature: 5~40 ℃ -20~70 ℃ Relative humidity: ≤ 80% ≤ 95%◆TemperatureChannel number: 2ch Measurement range: 25~45 ℃ Accuracy: ±0.1 ℃ Resolution: 0.1 ℃ response time: ≤ 3 min◆Power Requirements Input: 98-260 V AC, 50/60Hz
Consumption: ≤ 80VA